Figure 2-7 Sensor (Average Path) Functional Test Setup - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

2
Performance Tests
1 Connect the equipment as shown in

Figure 2-7 Sensor (Average Path) Functional Test Setup

2 Preset the instrument:
Press
3 Set the sensor mode to AVG only:
Press
• Highlight the Sensor Mode parameter using the up/down arrow keys.
• Press
• A pop- up window is displayed.
• Scroll through the Sensor Mode values using the up/down arrow keys.
Select a value of AVG only.
• Press
• Press
4 Zero the instrument:
Press
wait symbol to disappear.
5 Press
6 Connect the 30 dB attenuator as shown in
44
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Power Meter
Power Sensor
Power ref
CHANNEL A
Power sensor cable
Preset
then
.
Confirm
Local
.
Channel
Change
.
Enter
.
.
Done
Zero
,
Zero
. Wait approximately 25 seconds for the
Cal
Cal
to calibrate the power meter.
Figure 2-
7.
Figure 2-
8.
E4416A/E4417A Power Meters Service Guide