Figure 6-6 Troubleshooting Guide - Ttl Inputs (Continued) - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

No fault found with Channel A Input Zero/CAL
selection. Check user application/implementation.

Figure 6-6 Troubleshooting Guide - TTL Inputs (continued)

E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
A
Did power
meter channels
zero?
Yes
Apply TTL logic 1 level to RJ45 connector. pin 5 wrt
pin 8. Apply TTL logic 0 pulse to RJ45 connector
pin 6 wrt pin 8. See User's Guide chapter 2 for timing.
Did Channel A
CAL?
Yes
Is the
meter single or
Single
dual channel?
Dual
Apply TTL logic 1 level to Rmt I/O connector pin 5 wrt pin 8.
Apply TTL logic 0 pulse to Rmt I/O connector pin 6 wrt pin 8.
See User's Guide chapter 2 for timing.
Did Channel B
CAL?
Yes
No fault found with Channel B TTL Input selection.
Check user application/implementation.
End
Troubleshooting
6
No
No
No
B
123