Figure 6-2 Troubleshooting Guide - Ttl Output 1 (Continued) - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

Select
Fail O/P High
DVM reading
Yes
0.0V+/-0.5V?
No Fault found with Upper
Window TTL Over Limit
detection circuit - check user
application/implementation

Figure 6-2 Troubleshooting Guide - TTL Output 1 (continued)

E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
A
Select
Preset/local
Confirm
Meas/Setup
Limits
Limits On
TTL Output
TTL Output On, Fail O/P Low
DVM reading
Yes
5.0V+/-0.5V?
No
A2 Processor Assy
Defective
End
Troubleshooting
No
Power down meter.
Disconnect ac power and battery if fitted.
Remove A2 Processor assy, as directed on page
5-9, to access the Comms Assembly connector.
Use DVM to check resistance between J4 pin 41
and Rmt I/O connector pin 3.
DVM reading
Yes
450Ω +/- 10%
A4 Comms Assy
6
No
Defective
119