Agilent Technologies E4416A Service Manual page 58

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

This specification includes the ±0.4% setting and also takes into account other
factors that will contribute to determining the accuracy after 2 years. The
complete breakdown of the ±0.5% specification is as follows:
• Test line limit: 0.1%.
• System uncertainty: 0.3%.
• 2 year oscillator drift: 0.04%.
• Delta environment: <0.06%.
The same equipment used to measure the accuracy of the power reference
oscillator can also be used to set it to the post- factory accuracy of ±0.4% (for
further information refer to
The following procedure ensures maximum accuracy in measuring power
reference oscillator output using Agilent test instruments of known capability.
Signal acquisition criteria may vary if alternative test instruments are used. In
such cases refer to the manufacturer's guidelines for further information.
The test is performed using Equation 2- 4:
Equation 2-4: Power Level Provided By The Reference Oscillator
P
meas
The components of this equation are defined as follows:
• V
is the voltage measured between V
0
and after the Agilent 432A has been zeroed.
• V
is the voltage measured between V
1
power is approximately 1 mW.
• V
comp
ground with power applied.
• R is the resistance of the mount resistor in the Agilent 432A power meter.
For further information refer to
• Calibration Factor is the value of the thermistor mount at 50 MHz
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Chapter
2
(
)
2V
V
V
+
V
comp
1
0
0
=
-------------------------------------------------------------------
(
4R CalibrationFactor
is the voltage between the Agilent 432A V
Performance Tests
3, "Adjustments").
2
V
1
)
and V
with no power applied
comp
RF
and V
with power applied. This
comp
RF
connector and chassis
comp
Appendix
A.
2
37

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4417a

Table of Contents