Zero Test; Specification; Description; Equipment - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

2
Performance Tests

Zero Test

Specification

Electrical Characteristics
Accuracy: Zero set (Digital settability of zero)
*This performance limit is determined by the zero set specification of the power sensor used in the
measurement plus the measurement noise. The range calibrator has a zero set specification of
±0.05 mW. The calibrator measurement noise specification is 110 nW at 16 averages. At 512
averages a noise multiplier of 0.24 is required, giving a measurement noise specification of
0.0264 mW (0.24x512).

Description

After the power meter is initially calibrated and zeroed, the change in the
digital readout is monitored. This test also takes drift and noise into account,
since drift, noise and zero readings cannot be separated.

Equipment

• Range Calibrator: Agilent 11683A
• Power Sensor Cable: Agilent 11730A
18
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Performance Limits
*
±0.0764 mW
E4416A/E4417A Power Meters Service Guide