Removing The A5 Daughter Or A6 Measurement Assemblies - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

Replaceable Parts
5

Removing the A5 Daughter or A6 Measurement Assemblies

1 Disconnect the flex circuit from measurement assembly.
Take care when disconnecting the flex circuit from the measurement assembly.
N O TE
2 Loosen release tabs on measurement assembly and then remove vibration
support bracket.
3 Slide out the measurement assembly from the side of the power meter.
4 The daughter assembly is removed vertically.
A5
A6
E4416A/E4417A Power Meters Service Guide
101