Equipment

• Power sensor: Agilent E9321A.
• 2 × Power sensor cable: Agilent E9288A.
• 30 dB reference attenuator: Agilent 11708A.
• Digital voltmeter (DVM): Agilent 3458A.
• Breakout box kit: Agilent E9288 opt. K01
• (comprises a breakout box (coupler) and a pair of interconnecting red/black
leads).

Test Setup

Power sensor
Power meter
Power ref
CHANNEL A
Power sensor
cable

Figure 2-2 Instrument Accuracy (Normal Path) Test Setup

22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Power sensor cable
Breakout box
red
black
Digital Voltmeter
E4416A/E4417A Power Meters Service Guide
+ input
− input