Trigger - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

• Channel Functions:
• Storage Registers:
• Predefined setups:

Trigger

• Sources:
• Time Resolution:
• Delay Range:
• Delay Resolution:
• Hold- off:
• Internal Trigger:
• Level Accuracy:
• Resolution:
• Latency:
• External Trigger
Range:
• Trigger Out:
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
A, B, A/B, B/A, A− B, B− A, and relative.
10 instrument states can be saved via Save/Recall
menu.
For common wireless standards (GSM900, EDGE,
NADC, iDEN, Bluetooth, IS- 95 CDMA, WCDMA
and cdma2000),predefined setups are provided.
Internal, External TTL, GPIB, and RS232/422.
50 ns.
±1.0 s.
50 ns for delays <±50 ms otherwise 200 ns.
μ
Range: 1
s to 400 ms
Resolution: 1% of selected value (minimum of
100ns).
Range: − 20 to +20 dBm
±0.5 dB
0.1 dB
500 ns ± 100 ns
Latency is defined as the delay between the
applied RF crossing the trigger level and the
power meter switching into the triggered state.
High >2.0 V, Low <0.8 V; BNC connector; rising or
falling edge triggered; input impedance >1 kΩ
Output provides TTL compatible levels (High >2.4
V, Low <0.4 V) and uses a BNC connector.
Specifications
1
9