Measurement Uncertainty Analysis - Power Reference Level Test

Combined Standard Uncertainty (uc(y))

Table C-2
Combined Standard Uncertainty
Symbol
Source of Uncertainty
U(V
)
Two Voltage measurements
0
on a
3458A DMM
U(V
)
Two Voltage measurements
1
on a
3458A DMM
U(V
)
Voltage measurement on
comp
3458A DMM
U(R)
4 wire Resistance
measurement on
3458A DMM
U(CF)
Calibration factor for
Thermistor
U(M
)
Mismatch between the
s
DUT and the Thermistor
U(Drift)
Drift of 8478B Thermistor
U(Rep)
Repeatability of the
complete measurement
U(c)
Combined Uncertainty
*
U
Expanded Uncertainty
* Using Agilent's standard 95.5% confidence the coverage factor is k = 2.00
Expanded Uncertainty
The expanded uncertainty is: U= 3.030 mW, as a percent of 1.000 mW: U =
0.303%
C-148
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Value
Probability
Divisor
Distribution
69.65 μW
Normal
2
69.45 μW
Normal
2
49.25 μW
Normal
2
0.0046 Ω
Normal
2
0.002
Normal
2
0.0012
U shaped
0.0006
Normal
2
0.057 μW
Normal
1
Normal
k = 2
Normal
E4416A/E4417A Power Meters Service Guide
Ci
Ui/μW
-04
2.129x10
V/Ω
0.007
-02
1.224x10
V/Ω
0.426
-02
− 1.245x10
V/Ω
0.307
-06
− 4.727x10
V/Ω
0.012
-04
− 9.696x10
V/Ω
1.070
-04
− 9.454x10
2
V/Ω
0.885
-03
1.00x10
V/Ω
0.6
1
0.057
1.515
3.030