Test Procedure 2; Equipment; Test Setup; Figure 2-5 Instrument Accuracy (Average Path) Test Setup - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

2
Performance Tests

Test Procedure 2

This test procedure makes use of a 11683A Range Calibrator fitted with Option H01.
N O TE
The measurement uncertainty of this test procedure is significantly lower than the
previously described method (refer to
Instrument Accuracy Test
As such, this test procedure should be used to ensure the power meter meets its
published specification.

Equipment

• Range Calibrator: Agilent 11683A Option H01
• Power Sensor Cable: Agilent E9288A
• DC Calibrator: Datron 4000A

Test Setup

CHANNEL A

Figure 2-5 Instrument Accuracy (Average Path) Test Setup

32
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
for a description of the measurement uncertainty analysis).
Power Meter
Power sensor cable
Appendix D,
Measurement Uncertainty Analysis -
DC Calibrator
Range Calibrator
E4416A/E4417A Power Meters Service Guide