Therefore to attain a power of 100 mW as shown in the
applying 14.4493 Volts, the actual amount of voltage required from the
Universal Source would be:
14.4493 × 1.075533 = 15.6 Volts
By using the DMM Uncertainties calculated previously the worst case voltage
and corresponding power errors can be calculated.
Power Setting (mW)
Applied Volts
0.001
0.14493 mV
0.01
1.4493 mV
0.1
14.493 mV
1
144.93 mV
10
1.4588 V
100
15.6 V
E4416A/E4417A Power Meters Service Guide
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Measurement Uncertainty Analysis - Instrument Accuracy Test
DMM Error on Range
Worst Case Voltage
Standard Uncertainty
Error on Range (%)
μ
(
V)
μ
<1.371
V
0.946%
μ
<1.371
V
0.0946%
μ
1.371
V
0.00946%
μ
6.09
V
0.00420%
μ
56.8
V
0.00389%
μ
148.8
V
0.00095%
D
Table D-
2, instead of
Worst Case Power
Error on Range
μ
0.00946
W
μ
0.00946
W
μ
0.00946
W
μ
0.042
W
μ
0.389
W
μ
0.95
W
D-157