EPAP Step
The size of each incremental adjustment of the EPAP.
Unit
cmH
Relax Time
The time a status of normal breathing shall be maintained before the algorithm
is allowed to decrease the actual applied EPAP.
Unit
Minute
4.2.2

Ramp Up

The ramp up setting defines a ramp time for increasing the airway pressures.
At the beginning of Ramp Up time, IPAP will start at 2 cmH
Ramp Pressure. If the Ramp Pressure was set below EPAP, the EPAP pressure will start
at the configured Ramp Pressure.
Unit
Min
Minutes
10
Off
4.2.3

Ramp Down

The ramp Down setting defines a ramp time for decreasing the airway pressures.
At the end of Ramp Down time, IPAP will end at 2 cmH
Ramp Pressure. If the Ramp Pressure was set below EPAP, the EPAP pressure will end
at the configured Ramp Pressure.
Unit
Min
Minutes
10
Off
56
Treatment Functions and Settings
User manual
Min
Max
O
0.5
2
2
Min
Max
2
12
Off
Max
Default
60
Off
Max
Default
60
Off
Defau-
lt
Resolution
1
0.5
Defau-
lt
Resolution
5
1 min
O above the configured
2
Resolution
10
O above the configured
2
Resolution
10
Tolerance
± (0.5 cmH
0
2
+ 5%)
Tolerance
5 %
Tolerance
5 %
Tolerance
5 %
Doc. 007229 D-3