Operation
Setting Vaporizer Concentration
Refer to the appropriate Instructions for Use for the
vaporizer being used. Vapor 2000 is shown and
described below.
For the Dräger Auto Exclusion system:
Close any open vaporizers.
Ensure that the vaporizer is properly seated.
1 With the handwheel set to »T« position, press
the button and engage the handwheel at »0«.
Wait five seconds for the pressure to balance.
2 Press the button and turn the handwheel coun-
terclockwise to set the required anesthetic gas
concentration.
Regularly check the filling level on the sight
glass. When reaching the minimum mark, fill the
vaporizer with anesthetic agent.
O
Flush
2
A manually operated O
the front of the machine. When actuated, the valve
delivers an unmetered flow of at least 35 L/min to
the breathing system and breathing bag while
bypassing the ventilator. The Fabius OS does not
need to be switched on in order to use the O
3 Press the O
flush button. Additional O
2
into the compact breathing system as long as
the button is pushed in. The flow control ele-
ments and the anesthetic agent vaporizer
(Vapor) are bypassed.
NOTE:
In ManSpont mode, pressure may rise rapidly up
to the setting of the APL valve.
72
flush valve is located on
2
2
flows
2
flush.
Instructions for Use Fabius OS SW 3.n
2
1
3