Setting Vaporizer Concentration; O 2 Flush - Dräger Fabius OS Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius OS:
Table of Contents

Advertisement

Operation

Setting Vaporizer Concentration

Refer to the appropriate Instructions for Use for the
vaporizer being used. Vapor 2000 is shown and
described below.
For the Dräger Auto Exclusion system:
Close any open vaporizers.
Ensure that the vaporizer is properly seated.
1 With the handwheel set to »T« position, press
the button and engage the handwheel at »0«.
Wait five seconds for the pressure to balance.
2 Press the button and turn the handwheel coun-
terclockwise to set the required anesthetic gas
concentration.
Regularly check the filling level on the sight
glass. When reaching the minimum mark, fill the
vaporizer with anesthetic agent.
O
Flush
2
A manually operated O
the front of the machine. When actuated, the valve
delivers an unmetered flow of at least 35 L/min to
the breathing system and breathing bag while
bypassing the ventilator. The Fabius OS does not
need to be switched on in order to use the O
3 Press the O
flush button. Additional O
2
into the compact breathing system as long as
the button is pushed in. The flow control ele-
ments and the anesthetic agent vaporizer
(Vapor) are bypassed.
NOTE:
In ManSpont mode, pressure may rise rapidly up
to the setting of the APL valve.
72
flush valve is located on
2
2
flows
2
flush.
Instructions for Use Fabius OS SW 3.n
2
1
3