O 2 Flush - Dräger Fabius Tiro Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

If the control dial is in position T:
1 Press the 0 key (A) and engage the control dial
(B) to position 0. To ensure pressure
equalization, wait at least 15 seconds.
2 Press the 0 key (A) and set the control dial (B)
counterclockwise to the desired anesthetic gas
concentration.
3 Check the filling level on the sight glass (C)
regularly. The filling level must be between
minimum and maximum.
4 When the refill mark (D) is reached, 250 mL
(normal anesthetic agent bottle) can be refilled.
5 Fill the vaporizer at the latest when the
minimum mark (E) is reached, refer to
instructions for use of the respective vaporizer.
O
flush
2
The O
flush is used for flushing and quickly filling
2
the breathing system and breathing bag with O
while bypassing the vaporizer.
During this an unmetered flow of at least 35 L/min
is given to the breathing system and the breathing
bag.
 Press the
(A) key. O
the key is held down.
Instructions for use Fabius Tiro SW 3.n
A
+
2
flows for as long as
2
Use of the O
flush can increase airway pressure
2
very quickly and abruptly change the gas
concentration.
NOTE
In ManSpont mode, the pressure can increase
quickly and thus trigger the APL valve.
Operation
89