Assembly; Before First Use; Connecting The Gas Supply - Dräger Primus Infinity Empowered Quick Reference Manual

Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Assembly

NOTE
The Primus Infinity Empowered Quick Reference Guide is not a replacement or substitute for the Instructions for
Use. Any use of the device requires full understanding and strict observation of the Instructions for Use.

Before first use

Charging the battery for emergency
operation
Primus Infinity Empowered has a built-in uninterruptible
power supply UPS which maintains the power supply for
at least 30 minutes (up to 90 minutes, depending on the
ventilation parameters) in the event of a mains power
failure, provided that the battery is charged.
Switching to battery power (UPS) takes place automat-
ically, and is indicated on the screen by the message:
POWER FAIL.
The battery recharges automatically when the worksta-
tion is plugged into the mains, but only up to a maximum
ambient temperature of 35 °C (95 °F).
The battery must be charged for 10 hours before using
the workstation for the first time:
Plug the mains power plug of the Primus
Infinity Empowered workstation into the mains sock-
et.
The mains voltage must correspond to that specified on
the rating plate on the back of the machine.
Quick Reference Guide Primus Infinity Empowered SW 4.5n
Primus
A
The green LED
(A) lights up.
Leave Primus Infinity Empowered connected to the
electrical power supply for 10 hours. It does not need to
be switched on.

Connecting the gas supply

N
O AIR
2
1
Screw on the compressed gas hoses of the central
gas supply (CS) for O
nections of the gas inlet block (A). The two outer
ports at the back are reserved for the reserve gas
cylinders.
A compressed air outlet for the optional secretion
aspiration and an O
2
tube are optionally available.
2
Plug the other end of the pressure hoses into the ter-
minal unit.
3
Make sure all supplies are connected correctly and
functioning properly.
Assembly
A
O
2
, AIR, N
O to the front con-
2
2
outlet for an external O
2
flow
13