Drager Medizintechnik
2.2
Test 002: Testing the Gas Supply
This test allows you to test the gas supply.
•
Press the Reset key to select Test 002.
•
Press the Info key and the Reset key simultaneously to activate Test 002.
Display:
Pressure high
The gas supply pressure is sufficient.
•
Disconnect the Oxylog 2000 from the gas supply.
• Set the NoAirMix/AirMix mode selector switch to the centre position in order to vent the
residual pressure.
Display:
The gas supply pressure is too low or the
device is not connected to a gas supply source.
Pressure low
•
If problems occur during this test, first calibrate the pressure sensors (Test 11 in the
Extended Service Mode ESM)
•
Press the Info key and the Reset key simultaneously to return to the Customer Service
Mode (CSM).
Exiting the Customer Service Mode (CSM)
• Switch the Oxylog 2000 off and on again to return to the Oxylog 2000 normal operating
mode.
DS-Modus 5503.160 Oxylog 2000
11/97
Service-Mode SW 2.xx
Page 11