System Setup; Curve Optical Tracking - Brainlab CURVE System And Technical User Manual

Dual navigation station
Hide thumbs Also See for CURVE:
Table of Contents

Advertisement

System Setup

2.2
System Setup
2.2.1

Curve Optical Tracking

General Information
The Curve system is an optical-tracking, touchscreen-operated planning and navigation system
designed for pre- and intraoperative use.
• Monitor Cart and Camera Cart are delivered pre-installed and ready for use.
• All system components are suitable for continuous use during surgical procedures.
• Only use system components indoors, in an operating room.
Sterile Field
The Curve system is unsterile.
Warning
Do not allow the camera, monitors or any other part of the Curve system to enter the sterile
field.
Interferences
If camera's infrared light interferes with other devices, reposition these devices and/or the camera
so that the interference is resolved.
The system produces electromagnetic fields that may interfere with other sensitive equipment,
and can itself be disturbed by other electromagnetic fields.
Electromagnetic Compatibility
Curve requires special precautions regarding electromagnetic compatibility (EMC) and must be
installed and put into service according to the EMC information provided in this guide.
Warning
For EMC reasons, do not use system components adjacent to or stacked on other
equipment. If it is unavoidable, verify that the Curve system operates normally.
MR Safety
Warning
The Curve system has not been tested in an MR environment.
How to Safely Position the System
Step
1.
2.
3.
20
Remove protection covers from display and camera before entering the OR.
Move equipment into the OR.
Angle the system so that:
System and Technical User Guide Rev. 1.0 Curve Dual Navigation Station Ver. 1.2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents