TRANSPORT
7.4.4

Long Term Storage

General Information
Before and after placing the system into storage for an extended period of time, steps must be
taken to ensure proper functioning.
Related Links
1.2 Legal Information on page 8
8.1 Cleaning Instructions on page 127
10.2 Recurrent Test Requirements on page 146
How to Store for 1-6 Months
Before storing system for one to six months:
Step
1.
2.
3.
4.
How to Retrieve Equipment Stored for 3 Months
Before using equipment that has been stored for approximately three months:
Step
1.
2.
3.
Storage for 6 Months or Longer
Before storing system for six months or longer:
Step
1.
2.
3.
4.
How to Retrieve Equipment Stored for 6 Months
Before using equipment that has been stored for six months or longer:
Step
1.
2.
System and Technical User Guide Rev. 1.0 Curve Dual Navigation Station Ver. 1.2
Make sure that the UPS battery is fully charged.
Unplug all external cabling and store it together with the equipment.
Cover camera and monitors with protection covers.
Store equipment in a dry place with a temperature between 21°C (70°F) to 30°C (86°F).
Clean the equipment.
Fully recharge UPS battery.
Start system and verify correct function before using it for surgery.
Contact Brainlab support to remove the UPS Battery from the equipment.
Unplug all external cabling and store it together with the equipment.
Cover camera and monitors with protection covers.
Store equipment in a dry place with a temperature between 21°C (70°F) to 30°C (86°F).
Clean the equipment.
Contact Brainlab support to install a new UPS battery.