Protection Cover for Camera

8.1.4
Protection Cover for Camera
Illustration
Do Not Use
• Acid solvents
• Caustic solvents with pH value higher than 9.5
How to Clean the Protection Covers
Step
1.
2.
3.
4.
NOTE: If a soiled cover cannot be sufficiently cleaned using the above steps, exchange the cover
for a new one.
Do not wash, soak, iron, disinfect or sterilize protection covers.
132
Remove protection cover from Camera Cart.
Use a surface disinfectant to clean the surface.
Follow disinfectant manufacturer's recommendations.
NOTE: Use only a moist cloth to clean surfaces. Do not allow the protection cover to be-
come wet or soaked with disinfectant.
Clean the outside surfaces of the protection cover. Do not clean surfaces on the inside of
the cover.
Allow the protection cover to fully dry before applying to the system.