Message Boxes And System Warnings - Brainlab ROBOTICS VERO Hardware User's Manual

Table of Contents

Advertisement

ROBOTICS VERO
2.3.3

Message Boxes And System Warnings

List of Message Boxes and System Warnings
Messages
Hardware User Guide Rev. 1.1 Robotics Vero Ver. 2.1
Explanation
Movement is not possible as the Auxiliary Power Supply
(Battery Charger) is connected to Robotics Vero.
Movement is not possible as Robotics Vero is locked man-
ually by the Service Lock.
Disengage the Service Lock bolts to un-lock the device
and click OK for confirmation.
Robotics Vero is not connected.
The communication-link to Robotics Vero could not be es-
tablished or was lost during operation. Ensure that Robot-
ics Vero is powered and click OK for confirmation.
The Robotics QA-process requires a homing procedure.
An internal system-balancing process requires a homing
procedure, e.g. after a Service Brake release was per-
formed.
Click Acknowledge to start the homing procedure.
Click Cancel if you do not want to proceed with this action
now.
NOTE: During the homing procedure Robotics Vero must
be mechanically unloaded. No patient may lie on the couch
top during the homing procedure.
A communication error occurred.
The communication-link to Robotics Vero was disturbed
during operation.
If the problem persists, there may be a temporary disturb-
ance, or a hardware failure may be present in the communi-
cation-link.
Click Acknowledge for confirmation.
The application will shut down now.
If necessary move Robotics Vero to a valid default position
for patient treatment.
The message box appears at shut down of the application
to guide you through the controlled shut down procedure.
Select Home Position or Center Position to move Robot-
ics Vero to a valid default position for further patient treat-
ment.
Click Cancel to skip final adjustment of Robotics Vero po-
sition.