COMPLIANCES AND SPECIFICATIONS
11.4.5

Tested Cables

Use of Specified Cables
Warning
The use of accessories and cables other than those specified (with the exception of cables
sold by Brainlab as replacement parts), may result in increased electromagnetic emissions
or decreased electromagnetic immunity of the equipment and result in improper operation.
For a navigation station the following shall apply: S-Video type cables provided by the user shall
be equipped with a ferrite on both ends (e.g., Würth Electronic Art. No. 742 711 31 or equivalent).
USB type cables provided by the user shall be equipped with a ferrite on both ends (e.g., Würth
Electronic Art. No. 742 712 22 or equivalent).
Cable Specifications
Curve system cables that have been tested for emission and immunity conformity:
Cable
Curve Camera Cart Cable
Power cable
Potential equalization cable
SDI
DVI
Display Port
Hospital Network ("Intraoper-
ative Data")
Microscope
EM Basestation Data Cable
EM Basestation Power Cable Provided by Brainlab
System and Technical User Guide Rev. 1.0 Curve Dual Navigation Station Ver. 1.2
Specification
Provided by Brainlab
Provided by Brainlab
Provided by Brainlab
1 x BNC coax cable, shielded, terminated, 75 Ohm
1 x DVI-D cable 24 AWG, shielded, terminated, 100 Ohm
1 x DisplayPort 1.2 cable 30 AWG , shielded, terminated, 100
Ohm
Provided by Brainlab
Provided by Brainlab
Provided by Brainlab
177