MAINTENANCE
12.2

Malfunctions and Return Instructions

12.2.1

Malfunctions

System Damage or Failure
Do not continue to use the system if:
• The mains power cable or plug is damaged or frayed
• Liquid has been spilled into the device
• The system does not operate normally when operating instructions are followed
• A cart has tipped over or cover has been damaged
• System components exhibit a distinct decrease in performance, indicating need for servicing
• Liquids leak from system
• System emits smoke
• The LEDs indicate an error
If a error message appears on the touchscreens or if the error LEDs are lit, contact Brainlab
support. Do not use the system.
Related Links
3.3.5 LED Indications on page 48
5.4 LEDs and Acoustic Signals on page 94
How to Respond to Damage or Failure
Step
1.
2.
3.
4.
System and Technical User Guide Rev. 1.0 Curve Dual Navigation Station Ver. 1.2
Turn off system.
Unplug system from wall outlet.
Contact Brainlab support.
Attach a notice such as "DO NOT USE" to equipment to prevent it from being used inad-
vertently.