5.1.3

Calibrating the Flow Rate

Calibrate the flow rate if the Operational Qualification or Performance
Qualification flow rate accuracy and precision test fails.
Items needed:
0.076-mm (0.003-in) ID yellow PEEK tubing (P/N 049715) to create
14 ± 2 MPa (2000 ± 300 psi) of backpressure (if needed)
ASTM Type I (18.2 megohm-cm) filtered and deionized water
Balance capable of weighing more than 10 g with 0.001 g readability
Tared beaker
To calibrate:
1. To open the Wellness panel, click Wellness on the Chromeleon
ePanel (see
2. Under Pump Flow Rate Calibration, click Calibration Procedure.
A panel with instructions and command buttons appears.
3. Follow the instructions on the panel to complete the pump flow rate
calibration procedure.
4. Click Log Calibration Value to record the new calibration value in
the Chromeleon audit trail. This value is stored as the current value.
Doc. 155028-01 08/20
Figure
2-10).
NOTE Enter the weight of the collected water in grams
(to the nearest milligram).
5 • Service