Dionex Easion Operator's Manual
Pour les concentrés acides (comme le régénérant anionique), versez
toujours le concentré dans de l'eau désionisée et non dans le
réservoir vide.
Gießen
Anionregenerenten) das Konzentrat immer in entionisiertes Wasser
und nicht in den leeren Behälter.
6. Fill the bottle almost to the top with deionized water, and then place the bottle
in the tray on the top of the Dionex Easion.
NOTE To avoid staining the Dionex Easion, be careful not to
7. Using the concentrate bottle, pour additional deionized water into the bottle
until it is completely filled to the top. If a few drops spill over, then it is full
enough.
The regenerant bottle must remain filled all the way to the top at all
times.
8. Insert the stopper assembly tubing into the bottle and hand-tighten the cap.
9. Invert the bottle three or four times to disperse the concentrate.
After the analysis begins, do not mix the contents of the regenerant
bottle.
10. Verify that the liquid lines from the
to the corresponding lines from the Dionex Easion (see
30
Sie
säurehaltige
spill TBAOH on the system.
Konzentrate
(beispielsweise
and
ELUENT
REGEN
den
bottles are connected
Figure
3-2).