Vermeiden Sie Seitwärtsbewegungen, wenn Sie den Pumpenkopg
vom Kolben lösen. Andernfalls kann der Kolben brechen.
7. Place the pump head (front end down) on a clean work surface.
8. The piston does not come off as part of the pump head assembly because it is
captured by a magnetic retention system. After removing the pump head, hold
the shaft of the piston (near the base), tilt the piston slightly, and pull the
piston away from the pump.
Refer to
pump head.
Backup Piston Seal
74
Figure 5-6
or
Figure 5-7
Piston Seal
(P/N 055870)
O-Ring
(P/N 014895)
Spacer with Drain
(P/N 067705)
(P/N 063382)
Piston
(P/N 052840)
Figure 5-6. Primary Pump Head
when disassembling and reassembling the
Outlet Check
Valve Assembly
(P/N 045721)
Primary Pump Head
(P/N 067702)
Inlet Check
Valve Assembly
(P/N 045722)
Seal Holder
(P/N 067706)
Doc. 155028-01 08/20