Thermo Scientific Dionex Easion Operator's Manual page 51

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

6. Leave the syringe in the port (this prevents the sample from siphoning
out of the loop before injection).
7. Use one of the following methods to inject the sample onto the
column.
Loading Samples with a Vacuum Syringe (Pull Method)
1. Verify that the sample loading port on the front of the Dionex Easion
is connected to sample port
Figure
2. Disconnect the waste line from port
attach a shorter line: 25 to 30 cm (10 to 12 in) of PEEK or PTFE
(polytetrafluoroethylene) tubing (see
Loading Port
(on front door)
Figure 3-6. Loading Sample with a Vacuum Syringe (Pull Method)
3. Place the free end of the line into the sample.
Doc. 155028-01 08/20
Manually: Click Inject on the Chromeleon ePanel. For more
information about manual sample processing, see
Automatically: Include the Inject command in a Chromeleon
instrument method (see
about automatic sample processing, see
3-6).
Syringe
Sample
(5)
W
(6)
(1)
Injection Valve
Sample
3 • Operation and Maintenance
Section
3.11.6). For more information
Section
on the injection valve (see
S (5)
on the injection valve and
W (6)
Figure
3-6).
S
L
(4)
C
(3)
L
P
(2)
Section
3.11.2.
3.11.3.
43