Peak "Ghosting - Thermo Scientific Dionex Easion Operator's Manual

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

Flow rate through columns too high
Check the pump flow rate.
Clogged column bed supports
Refer to the column manual for troubleshooting procedures.
Contaminated columns
Clean the columns as instructed in the column manual.
Plugged Injection valve passages
Rebuild the injection valve (see
4.8

Peak "Ghosting"

Ghosting is the appearance of extraneous peaks in a chromatogram. These may be
late-eluting peaks from a previous injection, or they may result from a
contaminated, malfunctioning, or incorrectly installed injection valve. If these
peaks co-elute with peaks of interest, they will cause nonreproducible peak
heights and/or areas.
Insufficient time between sample injections
Wait until the previous sample has been completely eluted before making
another injection.
Insufficient flush between samples
Flush the sample loop with at least 10 loop volumes of deionized water or
sample between sample injections.
Incorrect or contaminated standards
Remake the standards.
Incorrect or contaminated eluent
Remake the eluent (see
Malfunctioning injection valve
Contact Technical Support for Dionex products for assistance.
Doc. 155028-01 08/20
Section
5.11).
Section
3.4).
4 • Troubleshooting
55