Wellness Panel Features - Thermo Scientific Dionex Easion Operator's Manual

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

Dionex Easion Operator's Manual
5.1.1

Wellness Panel Features

The Wellness panel provides the following features.
System Status
View and/or
Reset
Calibration
Parameters
Calibration
62
When the Dionex Easion is connected to Chromeleon, the
green Connected LED is displayed. When there is no
connection, the red Disconnected LED is displayed and all
other Wellness panel controls are disabled.
Click the pump On or Off button to turn the pump on or off,
respectively.
Click the Log Serial button to record the Dionex Easion
serial number in the Chromeleon audit trail.
The Calibration or Diagnostic box is green while a
calibration or diagnostic procedure is running.
The current electrochemical detector (ECD) conductivity,
pump Flow rate, and total system Pressure are displayed.
Click the Calibration Values button to open a window that
displays the current calibration values, the previous values,
and the factory-set values.
You can reset the current calibration values to the previous or
factory-set values from the window.
Pressure Transducer
Use the Calibrate Offset and Calibrate Slope buttons when
performing the pressure transducer calibration procedure (see
Section
5.1.2).
The Offset and Slope values and the date of the last
calibration are displayed.
Pump Flow Rate Calibration
Click the Calibration Procedure button to start the pump
flow rate calibration procedure (see
Click the Log button to record the calibration value in the
Chromeleon audit trail.
Section
5.1.3).
Doc. 155028-01 08/20