Incorrect suppressor operating conditions
Refer to the suppressor manual for troubleshooting procedures.
Cell above or below temperature
Contact Technical Support for Dionex products for assistance.
Damaged piston seal
Replace the piston seal (see
DCR waste backpressure tubing not installed (or installed incorrectly)
Verify that the DCR waste backpressure tubing is installed on the end of the
suppressor waste line. For more information, refer to the Dionex Easion
installation instructions.
Doc. 155028-01 08/20
Section
5.8).
4 • Troubleshooting
59