Vacuum
Degasser
A.6 Injection Valve

Injection Valve

A.7 Suppressors
Chemical
Suppression
Displacement
Chemical
Regeneration
Suppression
Capacity
Void Volume
Sequential
Suppression for
Anions
Suppressor Wear
Parts
Doc. 155028-01 08/20
Optional with external vacuum degasser
Two-position, six-port, electrically-activated Rheodyne valve
2 mm and 4 mm anion and cation
2 mm and 4 mm anion and cation
Dionex ACRS 500 (4 mm): 150 eq/min
Dionex ACRS 500 (2 mm: 37.5 eq/min
Dionex CCRS 500 (4 mm): 150 eq/min
Dionex CCRS 500 (2 mm): 37.5 eq/min
Dionex ACRS 500 (4 mm): <50 L
Dionex ACRS 500 (2 mm: <15 L
Dionex CCRS 500 (4 mm): <50 L
Dionex CCRS 500 (2 mm): <15 L
With a Dionex CRS 500 and a Thermo Scientific
Carbonate Removal Device:
The Dionex CRD 300 is designed for use with carbonate eluent-
based chemistries.
The Dionex CRD 200 is designed for use with hydroxide- and
borate-based chemistries.
None; peristaltic pump and inline filters not required