Loading And Injecting Samples With A Syringe - Thermo Scientific Dionex Easion Operator's Manual

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

Dionex Easion Operator's Manual
4. If the setting is wrong, change the device name.

3.11.5 Loading and Injecting Samples with a Syringe

This section describes two methods for loading sample into the injection
valve sample loop with a syringe:
Loading Samples with a Syringe (Push Method)
1. Verify that the sample loading port on the front of the Dionex Easion
2. Fill a syringe with a calibration standard or sample.
3. Insert the syringe into the sample loading port on the front of the
4. Verify that the injection valve is in the Load position. If it is not, click
5. Overfill the sample loop with five sample loop volumes. Excess
42
a. Select the Pump_InjectValve name and press the F2 key. The
Device Configuration dialog box appears.
b. Select AS from the Controlled By list.
c. Click OK.
Loading sample with a syringe through the sample loading port on the
front of the Dionex Easion (push method)
Loading sample with a vacuum syringe through the sample loading
port on the front of the Dionex Easion (pull method)
NOTE For instructions on how to load and inject
samples with an autosampler, see
is connected to sample port
Figure
3-6).
Dionex Easion.
Load on the Chromeleon ePanel.
sample will exit through the injection valve waste line.
NOTE Remember that the line from the syringe to the
valve must be filled before the sample loop can
be filled.
Section
on the injection valve (see
S (5)
3.11.4.
Doc. 155028-01 08/20