Dionex Easion Operator's Manual
5. Disconnect the orange
black
CELL OUT
6. Use an Allen wrench to remove the two screws on the top of the cell housing.
7. Remove the cell by unplugging it from the component mounting panel.
8. Plug the new conductivity cell (P/N 067761) into its mounting location.
9. Replace the screws on the top of the cell housing and tighten.
10. Attach the orange
black
CELL OUT
11. Attach the
12. Close the Dionex Easion front door.
13. Reconnect the power cord and turn on the Dionex Easion power.
82
Connector to component
mounting panel
CELL OUT (orange)
to CELL OUT (black)
Figure 5-10. Conductivity Cell Assembly
CELL OUT
line.
line from the cell to the union attached to the
CELL OUT
line.
line to the
ELUENT OUT
ELUENT OUT
to suppressor
line from the union that connects it to the
port on the suppressor.
ELUENT OUT
Doc. 155028-01 08/20