Example Chromeleon Commands For Loading And; Injecting Samples - Thermo Scientific Dionex Easion Operator's Manual

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

Dionex Easion Operator's Manual
4. Verify that the injection valve is in the Load position. If it is not, click
5. Insert a syringe into the sample loading port on the Dionex Easion
6. Use one of the following methods to inject the sample onto the

3.11.6 Example Chromeleon Commands for Loading and

Injecting Samples

The following example shows commands for loading and injecting
samples using a Dionex AS-DV.
Example Dionex AS-DV Instrument Method Commands
Sampler.LoadPosition
DeliverSample
0.000
Autozero
Wait
Inject
ECD_1.AcqOn
14.000
ECD_1.AcqOff
44
Load on the Chromeleon ePanel (see
front door and pull out the plunger to draw the sample into the
injection valve. Draw at least five times the sample loop volume.
NOTE The line from the sample to port
injection valve must be filled before the sample
loop can be filled.
column.
Manually: Click Inject on the Chromeleon ePanel. For more
information about manual sample processing, see
Automatically: Include the Inject command in a Chromeleon
instrument method (see
about automatic sample processing, see
CycleTimeState
Figure
2-10).
W (6)
Section
3.11.6). For more information
Section
;Switch the valve to Load.
;Deliver sample to the valve.
;Zero the baseline.
;Wait for cycle time (if any).
;Switch the valve to Inject.
;Start data acquisition.
;Stop data acquisition.
on the
Section
3.11.2.
3.11.3.
Doc. 155028-01 08/20