Liquid Leaks - Thermo Scientific Dionex Easion Operator's Manual

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

Dionex Easion Operator's Manual
4.2

Liquid Leaks

Leaking fitting
Locate the source of the leak. Tighten or, if necessary, replace the liquid line
connection (see
of Dionex Liquid Line Fittings (Document No. 031432).
Broken liquid line
Replace the line and fittings (see
Blocked or improperly installed line
Check that the lines are not crimped or otherwise blocked. Also, if the
blocked line is a waste line, make sure it is not elevated at any point after it
exits the Dionex Easion. If a line is blocked, replace it (see
Loose pump check valve housing
Make sure the check valves are firmly seated in the pump head. If they are
not, tighten them carefully with an open-end wrench just until the leak stops.
Damaged pump piston seal
1. Replace the piston seal (see
2. If the problem persists, replace the piston (see
Pump head not tight
Carefully tighten the pump head mounting screws just until the leak stops.
DO NOT OVERTIGHTEN!
Leaking pressure transducer
Check that the liquid line connections into the pressure transducer are tight.
Refer to Installation of Dionex Liquid Line Fittings (Document No. 031432)
for tightening requirements. Replace any damaged fittings.
Leaking pump head waste valve
1. Make sure the waste valve is closed. To close the valve, turn the knob
clockwise, just until tight. DO NOT OVERTIGHTEN! Overtightening
may damage the valve and the pump head.
2. Inspect the pump head. If the waste valve is the source of the leak, replace
the waste valve seal (see
52
Section
5.2). For tightening requirements, refer to Installation
Section
Section
Section
5.2).
5.8).
Section
5.10).
Section
5.2).
5.9).
Doc. 155028-01 08/20

Advertisement

Table of Contents
loading

Table of Contents