Thermo Scientific Dionex Easion Operator's Manual page 92

Ion chromatography system
Hide thumbs Also See for Dionex Easion:
Table of Contents

Advertisement

Dionex Easion Operator's Manual
3. The fuse drawer is located above the
power switch (see
tab locks the fuse drawer in place. Using
a small screwdriver, press the tab in and
then up to release the fuse drawer.
4. Pull the fuse drawer out of the rear panel
and remove the old fuses.
5. Replace the two fuses with new fast-blow
IEC 127 fuses rated 3.15 A
(P/N 954745). Thermo Fisher Scientific
recommends always replacing both fuses.
6. Insert the fuse drawer into the rear panel and press until the drawer snaps into
place.
7. Reconnect the main power cord and turn on the power.
84
Figure
5-11). A small
Fuse
Drawer
Tab
Figure 5-11. Fuse Drawer
Doc. 155028-01 08/20