3.11.2 Manual Sample Processing

To manually process a sample, select operating parameters and
commands from the Chromeleon ePanel. Commands are executed as soon
as they are entered.
Summary of Manual Sample Processing
1. Complete the instructions in
prepare the Dionex Easion for operation and to prepare the sample for
processing.
2. Load the sample into the injection valve sample loop with either the
Dionex AS-DV (see
3. Click Autozero on the Chromeleon ePanel.
4. Switch the injection valve to the Inject position with either the
Dionex AS-DV (see
The signal plot is displayed on the Chromeleon ePanel.
5. Monitor the chromatogram. When sample data has been collected,
click Monitor Baseline on the Instrument toolbar above the
Chromeleon ePanel Set.
Saving Manual Data
Data from manual processing is saved in the manual sequence under the
instrument folder in the local data vault.
To save the data from a manual run:
1. Select the manual folder and select File > Save As.
2. Enter a new name for the sequence.
3. Select the Save raw data check box.
4. Click Save.