Recommended Separation Distances From Portable And Mobile High-Frequency Communication Equipment; Reduced Separation Distances To Portable And Mobile Radio Frequency Communication - Dräger Babytherm 8004 Instructions For Use Manual

Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Recommended separation distances from portable and mobile high-frequency
communication equipment
The safety clearances listed in the following comply
with IEC 60601-1-2.
Max. P
(W) 150 kHz to 2.5 GHz
EIRP
0.03
0.32 m (1.1 ft)
0.10
0.58 m (1.9 ft)
0.17
0.76 m (2.5 ft)
0.20
0.82 m (2.7 ft)
0.25
0.92 m (3.0 ft)
0.41
1.2 m (3.9 ft)
0.82
1.7 m (5.6 ft)
1.00
1.8 m (5.9 ft)
1.64
2.4 m (7.9 ft)
3.3
3.3 m (11 ft)
Reduced separation distances to portable and mobile radio frequency
communication devices
The following separation distances are based on
tests performed by Dräger to determine the
minimum separation distances that are absolutely
necessary. These reduced separation distances
only apply to mobile high-frequency communication
equipment that uses the listed standards.
Mobile radio frequency communication equipment using ...
GSM 850, GSM 900, RFID 868 MHz (limited to 2 W ERP)
GSM 1800, GSM 1900 (limited to 1 W ERP)
UMTS, DECT (limited to 0.25 W ERP)
Bluetooth, WLAN 2450, RFID 2450 (limited to 0.1 W ERP)
Instructions for use Babytherm 8004/8010
All other frequencies Examples
0.96 m (3.2 ft)
1.8 m (5.9 ft)
2.3 m (7.6 ft)
2.5 m (8.2 ft)
2.8 m (9.2 ft)
3.5 m (12 ft)
5 m (16 ft)
5.5 m (18 ft)
7.1 m (23 ft)
10.00 m (33 ft)
WLAN 5250/5775 (Europe)
WLAN 2440 (Europe)
Bluetooth, RFID 2.5 GHz
WLAN 5250 (not in Europe)
UMTS cellular phones
Cordless DECT telephones
RFID 13.56 MHz
WLAN 5600 (not in Europe)
GSM 1800/GSM 1900
GSM-900 cellular phones, RFID
868 MHz
Separation distance
2.60 m (8.53 ft)
1.83 m (6 ft)
0.91 m (3 ft)
0.58 m (1.9 ft)
Technical data
117

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babytherm 8010

Table of Contents