Care; Disassembly; Cleaning/Disinfecting/Sterilizing - Dräger Babytherm 8000 WB Operating Instructions Manual

Heated bassinet
Hide thumbs Also See for Babytherm 8000 WB:
Table of Contents

Advertisement

Care

Care
The Babytherm 8000 WB heated bassinet must be
thoroughly cleaned and disinfected after each patient, or
at least once a week. Perform any disinfection
procedures according to established hospital procedures
as well as to the following additional instructions.
WARNING !
Always follow accepted hospital procedures for
handling equipment contaminated with body
fluids.
Before any cleaning procedures, the bassinet must first
be partially disassembled. For the cleaning and
disinfecting of accessories, see their respective
Operating Instructions.

Disassembly

Switch off the device(s), pull power plug, and turn
off all pressurized oxygen and air if used.
WARNING !
Always disconnect power supply before
cleaning and disinfecting.
Remove any ancillary patient care equipment from
unit.
Remove mattress from bed.
Store the matress flat only.
CAUTION !
Do not fold gel mattress!
Take off flexible U-grommets.
30
Cleaning / Disinfecting / Sterilizing
CAUTION !
Certain components of the bassinet consist of ma-
terials that are sensitive to certain organic solvents
sometimes used for cleaning and disinfecting (e. g.,
alcohols, phenols, halogen releasing com-pounds,
oxygen releasing compounds, strong organic acids,
etc.). Exposure to such substances may cause
damage that is not always immediately recognized.
Sterilization of the bassinet itself with ethylene oxide
(EtO) is also not recommended.
To prevent any damage, we recommend that only de-
tergents and disinfectants are used that are compatible
with the materials used in the bassinet and its compo-
nents and accessories, e. g. surface disinfectants on
the basis of
aldehydes, or
quarternary ammonium compounds
for disinfection procedures
Ensure that all disinfectants are registered with the U.S.
Environmental Protection Agency for use as intended.
Always follow the instruction labels specifically with
respect to prescribed concentrations and the necessary
exposure times.
Disassembly
Cleaning / Disinfecting /
Sterilizing

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fr 00 050

Table of Contents