Recommendations
on settings
Extreme settings above a defined threshold value or
inoperable settings are detected by the machine and
displayed es a message on the screen. At the same time
the green LEDs (operation indicators) flash on the rotary
knobs for setting.
Extreme settings occur if set v&es
exceed the following
threshold values.
In such cases Evita 2 limits the operative value to that
of the threshold value. The extreme setting does not
come into operation until the we&/check-
button
has been pressed.
pnlu :
5Ombar
I
pi"* :
30 mbar and 50 mbar
I
PEEP :
20 mbar
CPAP :
20 mbar
ASB :
30 mbar and 50 mbar
TI
:TE :
greater than 1 : 1
Or
smaller than 1 : 3
_