Monitoring
Configuring During Operation

Setting Monitoring Functions

Select »settings« column by turning dial knob,
press to confirm.
Display (example):
Setting volume of the pulse tone
0 = off
9 = maximum volume
Select the line »pulse tone« by turning dial knob,
press to confirm. The cursor automatically jumps to
the last value selected (example: 5).
Adjust value with dial knob, press to confirm.
Setting audible alarm volume
1 = minimum volume
9 = maximum volume
Select the line »alarm sound« by turning dial knob,
press to confirm. The cursor automatically jumps to
the last value selected (example: 4).
Adjust value with dial knob, press to confirm.
WARNING !
Always set audible alarm volume to a level
that is sufficiently loud with respect to the
general noise level of the surroundings.
NOTE: Local standards require a volume of at least
45 dB(A) for audible alarms. Therefore, settings
1 through 4 are configured to deliver this audio level.
Selecting the scale
For scaling volumeter function and trend trace.
Select the line »scale« with dial knob,
press to confirm.
Select »adult« or »child« with dial knob,
press to confirm.
78
CMV
MAC-config.
alarms
calibration
PAW
O
MAC display
default
40
2
21 Vol.%
on
off
CO
on
off
2
gas sensor
MAC correction
SpO
on
off
2
on
off
20
age
36
HLM on
off
FiCO
5
2
0
Fresh Gas O + N O
2
2
P
V
max
T
O %
L/min
cmH O
mL
2
2
50
1.00
25
600 10
settings
-Sensor
pulse tone
0 1 2 3 4
5
6 7 8 9
alarm sound
1 2 3
4
5 6 7 8 9
scale
adult
child
parameters
record
activate monitor defaults
T
:T
Freq.
T :T
PEEP
IP I
I
E
bpm
%
cmH O
2
1:2.0
10
0

Advertisement

Table of Contents
loading

Table of Contents