B Specifications; Environmental Specifications; Electrical Specifications - Waters ACQUITY Refractive Index Detector Overview And Maintenance Manual

Table of Contents

Advertisement

B Specifications

Environmental specifications

The following table lists the environmental specifications for the ACQUITY
refractive index detector.
Environmental specifications:
Attribute
Operating temperature
Operating humidity
Shipping and storage temperature
Shipping and storage humidity

Electrical specifications

The following table lists the electrical specifications for the ACQUITY
refractive index detector.
Electrical specifications:
Attribute
Protection class
Overvoltage category
Pollution degree
Moisture protection
Line voltages, nominal
Voltage range
Frequency
Maximum power draw
Fusing
a. Protection Class I
electrical shock. Class I identifies a single level of insulation between live parts (wires) and
exposed conductive parts (metal panels), in which the exposed conductive parts are connected
to a grounding system. In turn, this grounding system is connected to the third pin (ground
pin) on the electrical power cord plug.
94
October 13, 2014, 715003547 Rev. C
a
b
c
d
The insulating scheme used in the instrument to protect from
Specification
15 to 40 °C (59 to 104 °F)
20 to 80%, noncondensing
−30 to 60 °C (−40 to 140 °F)
10 to 85%, noncondensing
Specification
Class I
II
2
Normal (IPXO)
Grounded ac
100 to 240 Vac
50 to 60 Hz
145 VA
3.15 A, fast acting

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents