Waters ACQUITY Refractive Index Detector Overview And Maintenance Manual page 76

Table of Contents

Advertisement

4 Maintenance Procedures
Caution:
of a system module, do not disconnect an electrical assembly from
the module while the module remains connected to the ac supply
source.
Follow this procedure to completely interrupt power to the module:
Afterward, wait 10 seconds before disconnecting any assembly
from the module.
3.
Press down on the tab, to detach the leak sensor connector from the
front of the device.
Press down on
tab to release
connector
4.
Remove the leak sensor from its reservoir, holding it by its serrations
and pulling upward.
Leak sensor:
5.
Unpack the new leak sensor.
76
October 13, 2014, 715003547 Rev. C
To avoid damaging the electronic components or circuitry
1.
Set the module's power switch to Off.
2.
Disconnect the module's electrical supply cord from the ac
source.
Leak sensor connector
Serrations