Range And Attenuation; Features - Waters ACQUITY Refractive Index Detector Overview And Maintenance Manual

Table of Contents

Advertisement

1 The ACQUITY Refractive Index Detector

Range and attenuation

The total refractive index range of the detector is 1.00 to 1.75. However, for
optimum performance, the usable solvent refractive index range is 1.24 to
1.60. The measurement range of the instrument is 7 × 10
refractive index units full scale (RIU mode).

Features

Detector features:
Feature
Single
channel/additional
channel to monitor
temperature
Thermal stability
Single, programmable
contact closure output
Refractive index units
Purge
Long-life LED lamp
Recycle valve
24
October 13, 2014, 715003547 Rev. C
Description
The ability to monitor refractive index and flow
cell temperature.
A countercurrent heat exchanger design and
temperature-controlled, low-volume cell permit
stable operation.
A configurable switch accommodates a maximum
+30 V and 1 A. The switch can trigger fraction
collectors and other external and it can actuate
based on time and refractive index threshold
criteria.
The ability to provide measured refractive index
signals to nano-RIU sensitivity.
Programmable, for purging the reference flow cell
before performing a measurement.
For information about purging the
See also:
detector, consult the ACQUITY Console online
Help.
An 870-nm, long-life LED lamp.
A preconfigured diverter valve recycles solvent and
can be programmed to reduce solvent
consumption. The valve can be set to the waste or
recycle position via the console.
For information about recycling solvent,
See also:
consult the ACQUITY Console online Help.
to 5 × 10
–9
–4

Hide quick links:

Advertisement

Table of Contents
loading