Installing the leak sensor
Warning:
including inhalation, always wear chemical-resistant, powder-free
gloves and observe Good Laboratory Practice when you handle them.
See the Material Safety Data Sheets for the solvents you use.
Warning:
solvents, flush the detector with an appropriate miscible solvent prior to
performing this procedure.
Warning:
compounds that are toxic, wear clean, chemical-resistant,
powder-free gloves when performing this procedure.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Leak sensor
To install the leak sensor:
Caution:
of a system module, do not disconnect an electrical assembly from
the module while the module remains connected to the ac supply
source.
Follow this procedure to completely interrupt power to the module:
Afterward, wait 10 seconds before disconnecting any assembly
from the module.
1.
Power-off the detector.
2.
Open the detector door, gently pulling its right edge toward you.
3.
Carefully unpack the new leak sensor.
To avoid the harmful effects of personal contact with solvents,
To avoid the harmful effects of personal contact with leaking
To avoid personal contamination with biohazards or
To avoid damaging the electronic components or circuitry
1.
Set the module's power switch to Off.
2.
Disconnect the module's electrical supply cord from the ac
source.
Installing the leak sensor
October 13, 2014, 715003547 Rev. C
39