Input/Output Specifications; Performance Specifications - Waters ACQUITY Refractive Index Detector Overview And Maintenance Manual

Table of Contents

Advertisement

b. Overvoltage Category II
a local level such as an electrical wall outlet.
c. Pollution Degree 2
reduction of dielectric strength or surface resistivity. Degree 2 refers only to normally
nonconductive pollution. Occasionally, however, expect a temporary conductivity caused by
condensation.
d. Moisture Protection – Normal (IPXO) – IPXO means that no Ingress Protection against
any type of dripping or sprayed water exists. The "X" is a placeholder that identifies
protection against dust, if applicable.

Input/output specifications

The following table lists the input/output specifications for the ACQUITY
refractive index detector.
Input/output specifications:
Attribute
Analog output
Input
Output switch

Performance specifications

The following table lists the performance specifications for the ACQUITY
refractive index detector.
Performance specifications:
Item
Flow cell volume
Internal fluidic volume
Flow cell pressure rating
Flow cell material
Pertains to instruments that receive their electrical power from
A measure of pollution on electrical circuits that can produce a
Specification
10, 20, 40, or 80 points/sec, 18 bit
Range: -2.0 to 2.0 V
Maximum input voltage: 30 VDC
Logic "1": >3.3 VDC
Logic "0": <1.7 VDC
Rated current: 0.5 A max
Rated voltage: 30 VDC max
Contact resistance: 0.2 ohms
October 13, 2014, 715003547 Rev. C
Input/output specifications
Specification
1.3 μL
<15 μL
≤1000 kPa (10.0 bar, 145 psi)
Fused quartz
95

Hide quick links: