Appendix B Specifications and Layout
B
QuantStudio
3D Digital PCR Instrument layout and connections
QuantStudio

Instrument clearances

80
3D Digital PCR Instrument layout and connections
The QuantStudio
3D Digital PCR Instrument and the ProFlex
(or Dual Flat Block GeneAmp
benches or collocated as shown in the following figure. If you place the units
adjacent to one another, provide at least 20 cm (8 inches) clearance between them to
allow for heat dissipation.
1
QuantStudio
3D Digital PCR Instrument
1
ProFlex
2 × Flat PCR System
2
The instruments require the following additional clearances:
IMPORTANT! For safety, the outlet used to power the instruments must be
accessible at all times.
Component
QuantStudio
3D
Digital PCR Instrument
QuantStudio
3D
Digital PCR Chip Loader
ProFlex
2 × Flat PCR
System
Dual Flat Block GeneAmp
PCR System 9700
[1]
With the heated cover open.
[2]
Ensures adequate airflow and cooling.
PCR System 9700) can be placed on separate
ProFlex
PCR System
2
Top
Front
30.48 cm
30.48 cm
(12 in.)
(12 in.)
25.4 cm
15.2 cm
(10 in.)
(6 in.)
15.5 cm
[1]
(6.1 in.)
20.38 cm
(8 in.)
QuantStudio
2 × Flat PCR System
15.2 cm
(6 in)
20 cm
(8 in)
5 cm
(2 in)
Sides
15.2 cm
(6 in.)
15.2 cm
(6 in.)
20.38 cm
(8 in.)
3D Digital PCR System User Guide
CAUTION HOT
ProFlex
PCR System
Back
15.2 cm
(6 in.)
25.4 cm
(10 in.)
23.0 cm
[2]
(9 in.)
20.38 cm
(8 in.)