Observation
QuantStudio
™
3D Digital PCR System User Guide
Troubleshooting chip images using the Chip View
Possible Cause
Excess template was loaded
onto the chip.
Chapter 6 Troubleshooting
Action
™
In the AnalysisSuite
Software, check
the copies per reaction and dilute the
sample accordingly so that it is within
the digital range.
6