Calibrate the touchscreen
QuantStudio
™
3D Digital PCR System User Guide
3. Remove each fuse from its fuse holder and inspect it for damage. Carbon
typically coats the inside of failed fuses.
4. Replace each failed fuse with a 1.6A, Time-Lag T, 250VAC, 5 × 20-mm Fuse.
Note: The voltage and amperage ratings are on the fuse holder.
5. Install the fuse holder.
6. Plug in, then power on the QuantStudio
installation is successful if the instrument powers on.
Note: Fuse failure can result from fluctuations in the supplied power to the
™
QuantStudio
3D Digital PCR Instrument. To prevent further failures, consider
installing an electrical protective device, such as a UPS or surge protector.
Calibrate the instrument touchscreen periodically as part of the routine maintenance
of your instrument, if you observe a decrease in screen responsiveness over time, or
if prompted after updating the instrument firmware.
1. From the Main Menu of the touchscreen, touch
Service.
2. In the Maintenance & Service screen, touch Screen Calibration.
Appendix C Maintenance
Calibrate the touchscreen
™
3D Digital PCR Instrument. The
, then touch Maintenance &
C
91