Appendix A Install the QuantStudio
A
Install the GeneAmp
78
3D Digital PCR System
PCR System 9700
d. Configure the Security Code screen according to your needs.
User Name: hank
Press PIN # to create a #. Then you set
protection to Locked to prevent methods
from being overwritten or deleted.
Accept
Name
F1
F2
do not want to protect your methods
want to protect your methods
want to return to the Main menu
without adding the new user name
8. Program the GeneAmp
chips.
Table 5
GeneAmp
Stage 1
96.0°C
56.0°C
10 min
2 min
1x (Hold)
39x (Cycles)
[1]
The reaction volume on the instrument display does not refer to the reaction volume on the chip and
should not be changed.
1 Hold
2 Tmp 39 Cycles
96.0
10:00
60.0
2:00
Start
Method: exp 001
F1
F2
Note: For information on programming the GeneAmp
the GeneAmp
PCR System 9700 Base Module User Manual
(Pub. No. 4303481).
The installation of the GeneAmp
components are ready for use.
PIN number: None
Protection: Unlocked
PIN#
Cancel
F3
F4
F5
If you...
PCR System 9700 with the method for thermal cycling
PCR System 9700 PCR Method
PCR protocol
Stage 2
Stage 3
98.0°C
60.0°C
30 sec
2 min
1x (Hold)
2 Holds
98.0
60.0
0:30
2:00
10.0
Return
F3
F4
F5
PCR System 9700 is complete and the
QuantStudio
Then...
press Accept (F1) again.
see the GeneAmp
PCR System 9700
Base Module User Manual
(Pub. No. 4303481) for instructions on
securing your methods.
press Cancel (F5).
Run
speed
10.0°C
Standard
PCR System 9700, see
3D Digital PCR System User Guide
Reaction
volume
[1]
20 μL