Appendix A Install the QuantStudio
A
Install the GeneAmp
76
3D Digital PCR System
PCR System 9700
2. When the thermal cycler displays the main menu, confirm that the version
number displayed on the screen is 3.12 or later. If the version number is earlier
than 3.12, contact technical support to upgrade the instrument firmware before
continuing.
IMPORTANT! The GeneAmp
version 3.12 or later to thermal cycle the chips.
08:00 AM
4/25/13
GeneAmp® PCR System 9700
Version: 3.12
Name:tc001 User:<<pe>>
Run
Create
F1
F2
v3.12 or later
1
3. From the Main menu, press Util (F4).
08:00 AM
4/25/13
GeneAmp® PCR System 9700
Version: 3.12
Name:tc001 User:<<pe>>
Run
Create
F1
F2
4. From the Utilities screen, press Config (F3).
Utilities
Diag
- Instrument diagnostics
TmCalc - Calculates melting temp
Config - Instrument configuration
Diag
TmCalc
F1
F2
5. From the Instrument Configuration screen, set the instrument time:
a. Use the Circular Key Pad to select the Time field.
b. Press the 24 Hr (F2) or PM (AM) (F3) soft keys until the format you want for
the current time displays in the Time field.
c. Use the numeric keys to type in the hours followed by minutes.
d. Press Accept (F1) when your entries are complete.
Instrument Configuration
Time:
Date:
Run Time Printer: Off
Run Time Beep:
Accept
24Hr
F1
F2
Note: CE clears an entry.
6. From the Configuration screen, set the instrument date:
a. Use the circular key to select the Date field.
PCR System 9700 must be running firmware
25.0°C
Edit
Util
User
F3
F4
F5
25.0°C
Edit
Util
User
F3
F4
F5
Config
More
Exit
F3
F4
F5
11:30 AM
01/25/00 M/D/Y
Off
PM
More
Cancel
F3
F4
QuantStudio
1
F5
3D Digital PCR System User Guide